| 1. | Answer 2 of 3 (a)-(c). For this question, you must supply
specific details (instruments, sampling techniques, detectors, etc.) (4
points each, 8 total)
There are several options for each situation, I've supplied what I think
is the simplest (but not necessarily the best) method. In each case
I'm looking for a comment on each of the following:
Also: somewhere you should make a comment about running standards and making a calibration line (all three of these are quantitative) |
||||||
| 1. (a) | A truck carrying several types of PCB waste has crashed into a river.
How would you determine (quantitatively) how much of each PCB is present
in the river water at a point downstream of the spill? Note: PCB
= “polychlorinated biphenyl” - i.e. they each contain 2 benzene rings.
PCB's are non-volatile.
sampling: already in water, can be injected directly into machine (presumably
dilute enough)
|
||||||
| 1. (b) |
(b) How would you quantitatively analyse each of
the several volatile hydrocarbons in a sample of gasoline? These
compounds are saturated (no double bonds), usually isomers of C7H16
and C8H18.
sampling: liquid, can probably be injected directly into machine (if
diluted, in lighter hydrocarbon, e.g., hexane)
|
||||||
| 1. (c) | A drug has 2 chiral forms, R and S. The R-isomer
is toxic. Unfortunately, pure S-isomer can racemise (isomerise
into the R form) over time. It is expected that a patient
has taken expired medicine that has recemised. How would you analyse
the patient's blood to determine how much of each isomer is present?
(The drug is a solid containing benzene rings.)
sampling: some comment on removing proteins and such from blood (don't
want to inject blood in directly)
|
||||||
| 2. | Answer 3 of the following 4 (a) to (d) (2 points each, 6 total): | ||||||
| 2. (a) | Explain how ion-exchange chromatography works.
- stationary phase a gel with exchangable ions
|
||||||
| 2. (b) | Fluorescence does not follow Beer's Law (A = ebc).
Instead, under conditions of constant light source power, the formula F
= 2.3K'ebc can be used.
Explain what this extra term K' represents.
- e represents amount of light analyte will
absorb (whether photon is absorbs)
|
||||||
| 2. (c) | A pyridine base has a distribution constant between ether and water
of 3.25. A water sample (40.0 mL) is 0.300 M in the acid. What
is the concentration remaining in the water after 5 extractions with ether
if each extraction uses a volume of 10.00 mL?
|
||||||
| 2. (d) | Explain why capillary electrophoresis is both faster and has better
resolution than traditional gel electrophoresis.
- both N (number of plates) and v (velocity of electroosmotic
flow) are directly proportional to V (voltage)
|
||||||
| 3. | An HPLC experiment has the following parameters:
Length 24.7 cm Flow rate 0.313 mL/min VM 1.37 mL VS 0.164 mL
There are a number of ways (different formulae) to do this, but basically the steps are: (i) calculate number of plates in column, (ii) calculate number of plates needed to get desired resolution, (iii) calculate length required for that many plates. The easiest method (I think) is shown below:
|
||||||
| 4. | Occasionally, the military has to clean up old ammunition dumps. At one such ammo dump, it is known that a barrel contains a toluene-based explosive or precursor, but it is not known which one. Infrared absorption was used to determine the identity of the unknown. | ||||||
| 4. (a) | The compound was initially a solid. The solid could be shock-sensitive
(i.e. explode if it is ground up). How could you safely prepare and
run the sample?
If you cannot grind the compound, that rules out both KBr discs and nujol mulls. Therefore, either of the solution techniques is acceptable (thin film by evaporation of solvent, or running the IR of the solution itself). |
||||||
| 4. (b) | The compound is the carboxylic acid, as evidenced by the broad O-H peak (3100) accompanied by the carbonyl peak (C=O at 1690). The other two peaks are likely a C-O and the O-H bend, further verifying the identity as the acid. |
Last updated, March 5, 2003