| 1. | An HPLC experiment has the following parameters: Length 15.0 cm Flow rate 0.5 mL/min VM 1.37 mL VS 0.164 mL
|
||||||
| 1. (a) | Calculate the resolution between unknown peaks A and B, and between
peaks B and C. (2 points)
|
||||||
| 1. (b) | If RS α (N)1/2 and N
α L, then RS α (L)1/2
|
||||||
| 2. | Answer 2 of the following 3 (a) to (c) (2 points each, 4 total): | ||||||
| 2. (a) | Explain electroosmotic flow.
- at "normal" pH, some SiOH groups on capillary walls are deprotonated
to Si-O- (0.5) |
||||||
| 2. (b) | Explain how an electron-capture detector works - 63Ni radioactive source decays by β-emission (emits a high-energy electron) (0.5) - these particles hit the carrier gas (N2), creates a bunch more, lower energy electrons (0.5) - electronegative groups on analyte molecules easily attract these electrons (0.5) - loss of electrons to absorption is detected as current loss (0.5) |
||||||
| 2. (c) | Describe how a light-scattering HPLC detector works.
- solution comes off column into a nebuliser, forms a fine spray
(0.5) |
||||||
| 3. | Answer the following (3 points each, 6 total): | ||||||
| 3. (a) | Samples for infrared spectroscopy are commonly prepared in one
of these three forms: (i) nujol mull, (ii) solution, (iii) KBr pressed pellet.
Compare the relative merits (advantages and disadvantages) of 2
of these 3 methods. (i) advantage - easy to prepare and use (0.5) disadvantages - not reproducible/quantitative (0.5) and several interfering peaks (0.5) (ii) advantage - no interfering peaks (0.5) disadvantages - not reproducible/quantitative (0.5), requires a hydraulic press (0.5) (iii) advantage - quantitative in concentration disadvantages - requires special cell, sample needs to be soluble, solvent may have interfering peaks (1) |
||||||
| 3. (b) | Describe the basis for size-exclusion chromatography, including
a description of why size-exclusion gels have an upper and lower mass limit. - gel contains little pores into which molecules can fit (0.5) - small molecules fit into holes and so go in and out of pores, increasing distance travelled and therefore retention time (1) - the larger the molecule, the fewer pores they will fit into, so pathlength and retention time decrease as size of molecule increases (0.5) - therefore, any molecules larger than the upper limit (size of the largest pore) of the gel will come straight off with no separation (0.5) - likewise, any molecule smaller than the smallest pore size will enter every pore in its path and will therefore have the same pathlength (and retention time) (0.5) |
||||||
| |
|||||||
| 4. |
Answer the following (2 points each, 6 total) |
||||||
| |
|||||||
| 4. (a) | A phenol has a distribution constant between hexane and water
of 6.25. A contaminated water sample (50.0 mL) is 0.0600 M
in the phenol. What is the concentration remaining in the water after
5 extractions with hexane if each extraction uses a volume of 15.00 mL? (2
points)
|
||||||
| |
|||||||
| 4. (b) |
Describe the injection system of EITHER
a GC OR an HPLC instrument. GC: - must have a rubber septum to reseal after injection to avoid sucking in air (0.5) - must be hot to volatilise entire sample simultaneously and quickly (therefore requires small sample size) (1) - must have a connection to allow addition of carrier gas (0.5) HPLC: - need a sample loading loop where sample is drawn into a reservoir (1) - there is a toggle between the reservoir and the column that is switched over to expose sample to the high pressure of the instrument (1) |
||||||
| |
|||||||
| 4. (c) |
Explain why a molecule's emission (fluorescence)
spectrum is always at longer wavelength than its absorption spectrum - light is absorbed from the ground state to an excited vibrational and electronic state; there are many vibrational states possible (0.67) - the excited vibrational states always relax very fast, therefore, the excited electron quickly loses a little energy in vibrational relaxation (0.66) - the molecule then re-emits a photon to relax to the ground state which is of lower energy because some energy has been lost to vibrational relaxation (0.66) |
||||||
| |
|||||||
| 5. | Answer 2 of 3 (a)-(c). For this question, you must
supply specific details (instrument, sampling technique, detector,
etc.) (5 points each, 10 total)
There are several options for each situation, I've supplied what I think is the simplest (but not necessarily the best) method. The approximate breakdown in marks for (b) and (c) is as follows: - type of instrument used (1) |
||||||
| 5. (a) | Infrared spectroscopy is often used to analyse emissions from
internal combustion engines (e.g., car engines). Without
resorting to chromatography, explain how you would use IR to determine the
amount of CO, CO2, and SO2 that might be present in
the exhaust. Do any of these peaks overlap? Why or why not?
- CO has a triple bond, CO2 has C=O double bonds, therefore
they have different k so will not overlap (1) |
||||||
| |
|||||||
| 5. (b) | The "high test" gasoline at a local station claims to have 92
octane rating (meaning it is 92% "isooctane"). How would you determine
the components (amount and identity) of this gasoline? For this question,
let's assume (it's a big assumption!) the following components could be in
gasoline, all of which have an octane rating of 0 except isooctane. [Followed
be 5 compounds with their boiling points] - instrument: gas chromatography - separation method: probably should use capillary coated with stationary phase (capillary GLC) - inert gas as mobile phase (pick one of He, Ar, or N2) - detector: for quantitation, probably need FID - could run standard solutions to determine location of peaks (identification) and detector response (quantitation) for each component - some comment on sampling technique - e.g., can be injected neat (without dilution) |
||||||
| |
|||||||
| 5. (c) | You are a customs office that is trying to determine if the suspect's
pills are percocet or a legal over-the-counter acetominophen pain-killer (without
the active percocet component, 14-hydroxydihydrocodeinone). [structures for
caffeine, acetominophen, and 14-hydroxydihydrocodeinone were included] - sample must be made into a solution - instrument: HPLC (1) - separation method: non polar stationary phase (reverse phase) probably best - some comments on solvent required (say, a mixed water/acetonitrile or water/methanol mobile phase) - detector: any of the following will work - IR, UV-vis absorption, light scattering, mass spec (but must pick one!) - some comment on identifying the peaks required (e.g., running standards) |