| Text sections: | Supplementary Material (available through the Bookstore) |
| Questions: | Pick up at my office (on the counter across from my office door) |
Topics covered:
(note: figure numbers are from the supplementary material, not the
textbook)
| 1. | Basic Instrument - source, mass analyser, detector
- mass analyser serves same purpose as column in chromatography, separates be mass, sensitivity +/- 1 amu - actually separates by mass: charge ratio (m/z), ions positively charged Figure: 11-1 |
| 2. | Isotopes - sensitivity of +/- 1 amu means isotopes
are differentiated
- recall isotopes: same element (number of protons, electrons), different number of neutrons - nominal mass: proton = 1 amu, neutron = 1 amu, electron = 0 amu - in amu of isotope is sum of protons and neutrons - low resolution MS = differentiate molecules/ions by nominal mass only - high resolution MS = to higher number of sf, there is a difference in proton and neutron mass, and can see effect of binding energy, therefore can determine molecular formula exactly - natural abundance: some elements contain several naturally occuring isotopes, MS will detect characteristic isotope pattern - a statistical distribution of isotopes will exist if multiple atoms of such an element |
| 3. | Atomic Mass Spectroscopy - ICP-MS
- use inductively-coupled plasma source (ICP) - plasma breaks everything into component elements and ionises the atoms - skim plasma and run through mass analyser to separate atoms by mass Figure: 11-12 |
| 4. | Molecular Mass Spectroscopy Sources
(a) Hard sources - large amount of fragmentation (i) electron impact - gas-phase analyte molecules pass through an electron stream (ii) chemical ionisation - carrier gas added, ionised by electrons; carrier gas ions then ionise analyte (b) Soft sources - little fragmentation (i) fast-atom bombardment (FAB) - analyte molecules bombarded by high-speed atoms - analyte is in solid form - atoms bombard solid (ii) matrix-assisted laser desorption/ionisation (MALDI) - analyte mixed with absorbing species (matrix) - laser hits matrix, excites absorbing species, energy transfer to analyte to ionise into MS (iii) electrospray - analyte in solution is aspirated into MS, passes through voltage to ionise (iv) field ionisation - gas-phase analyte molecules pass through high voltage to ionise Figure: 20-3 |
| 5. | Mass analysers
(a) quadrupole - two positive and 2 negative poles, all hooked up to voltage that can cause polarity change - positive poles are low mass filters - all ions repelled - when polarity changed to negative, ions attracted, light ions crash and are eliminated - negative poles high mass filter - all ions attracted - when polarity changed, ions repelled, only light ions repelled enough to avoid crashing (b) magnetic sector analysers - ions accelerated through voltage, then sent through curved magnetic field - ions that are too light curve too fast and crash into inside wall - ions that are too heavy curve too slow and crash into far wall - by changing accelerating voltage, can control speed and therefore mass of ion that makes it through - double focussing = machine with two such curves for higher resolution where B = magnetic field of magnet r = radius of curve e = charge on electron (constant) V = accelerating voltage - the lighter the atom, the faster it's moving, the faster it reaches the detector - resolution = m/Dm
(~102 for low res, ~104 for high res)
Figures: 11-4, 11-6, 11-8, 11-9, 20-12, 20-13 |
| 6. | Quantitative analysis - usually involves some sort of standard
- can normally assume that detector has linear response over entire mass range (peak height quantitative) (a) calibration line with series of known concentrations of analyte (b) "spiking" with a non-interfering element (In or Rh, which have low natural abundancies) (c) addition of isotopically-labelled standard of analyte - 14C dating = quantitative ratio of 14C/12C |
| 7. | Qualitative analysis - identification of unknowns and use in
synthesis
(a) following reaction or identifying product of a reaction - with high res and isotopic analysis, able to determine molecular formula (b) fragmentation pattern - used to identify chemical units (e.g., amine, phenyl, carbonyl, etc.) - molecules break up through their weakest bonds, fragments indicate original connectivity of parts - complications - fragments could re-arrange (at low m/z), secondary fragmentations less predictable - limitations - difficult to determine certain types of isomers (optical; cis / trans; o-, m-, p-benzene) |